<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>40526</ErrorNumber>
<ErrorFamily>Media Handler</ErrorFamily>
<ErrorMessage>
MDH: Loadtable unexpectedly detected media in the center of the loadtable while staging media
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The loadtable attempted to stage the media from the shuttle onto the mainscan stage pins
however media was detected in the middle of table when it was a size that it was not expected
to be found in the middle of the table,
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Check the device to ensure media is not caught or blocked
</text></DescriptionText>
</OperatorAction>
<Availability>
<DescriptionText><text>
This error is reported.
</text></DescriptionText>
</Availability>
<Cause>
Faulty hardware.
</Cause>
<Cause>
Faulty sensors (LT_SensorSubscanStageCenterPlateDetect)
</Cause>
<LastModified>
03-Dec-2007
</LastModified>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 1 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 1 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllParameters.htm</LinkAddress><LinkTitle>Firmware 1 Parameter Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Thu Jul 17 07:23:28 2008 PST